MOOC: Validation of liquid chromatography mass spectrometry (LC-MS) methods (analytical chemistry) course

Self test 5.2B

1.

You are validating a LC/MS analysis of L-PFOS in ground water and for trueness evaluation, you are analyzing a commercial CRM sample.     

image001.png

You are using matrix matched calibration.            

Your analysis data is:      

Number of result

C (ng/L)

1

13.96

2

14.00

3

12.51

2.

Calculate bias (unit: ng/L):

3.

Relative bias (unit: %):

4.

For evalution of your bias, calculate zeta-score.

[accordion][acctitle]Clue 1[/acctitle]For u(lab) use standard deviation of your parallel measurements[/accordion]

[accordion][acctitle]Clue 2[/acctitle]For u(ref) use expanded uncertainty in the certificate divided by 2 (k = 2)[/accordion]

5.

Result is:

[accordion][acctitle]Clue[/acctitle]∣z∣≤ 2 means satisfactory results; 2<∣z∣≤3 means questionable results and ∣z∣>3 means unsatisfactory result.[/accordion]

6.

Possible reasons for unsatisfactory results (click all that apply):

7.

Method bias is:

8.

Could this CRM be used for evaluating L-PFOS method bias in the case of sea water?