Validation of liquid chromatography mass spectrometry (LC-MS) methods

Self test 7

1. What types of performance characteristics are described? Please choose correct characteristic.

Description Characteristic
Mean value of 6 measurement results of the same sample was 5.2 mg/L, reference value was 5.0 mg/L.  
Standard deviation of the mean of 6 titration results of the same sample was 0.02 mL.  
Repeatability of the method obtained as the pooled standard deviation from 127 measurement results.  
Measurement uncertainty was estimated as ±0.2 mg/L, k = 2.  
Relative expanded uncertainty of the method is in the range of 0.9 – 1.1%, k = 2, depending on measurement conditions.  
Analyte loss occurs during sample pre-treatment, causing systematically lower results. Lowering of the results has been estimated to be on an average 5%.  
2. Which of the following is true?
3. Which of the following is true?

4. In an interlaboratory comparison, Lab’s result is 12.6 mg/kg with combined standard uncertainty 1.1 mg/kg. The independent reference value is 13.8 mg/kg with combined standard uncertainty 0.5 mg/kg. Which of these – zeta-score, En number, z-score – can be calculated from these data? Please calculate them and present the answers with 2 decimals. If cannot calculate, please type N/A.

Answer
zeta-score
En number
z-score

Is the lab result in agreement with the reference value?


 