Validation of liquid chromatography mass spectrometry (LC-MS) methods

Self test 5.2B

You are validating a LC/MS analysis of L-PFOS in ground water and for trueness evaluation, you are analyzing a commercial CRM sample.

CoA 

You are using matrix matched calibration.

Your analysis data is:
Number of result

C (ng/L)

1

13.96

2

14.00

3

12.51

1. Calculate bias (unit: ng/L):

2. Relative bias (unit: %):

3. For evalution of your bias, calculate zeta-score.

4. Result is:

5.

Possible reasons for unsatisfactory results (click all that apply):

6.

Method bias is:

7.

Could this CRM be used for evaluating L-PFOS method bias in the case of sea water?

8. Please indicate which of the main problems from the given list can be expected to occur when carrying out spiking in the following analytical situations. Please make a choice in the following table.

Analytical situation Analyte is unstable Difficult to disperse the analyte
in the matrix
Analyte encompasses a number of
different individual compounds
Determining meropenem antibiotic in blood plasma
Determination of amino acids in tea
Determination of residues of explosives in soil
Determining anthocyanins in bilberries
Determining ascorbic acid in apples