Calibration
Calibration of anemometers using Rotary Jib Standard (PNE) and Wind Tunnel Standard (TTE)
|
price according to agreement |
We carry out calibrations of anemometers using two standards. For lower air flow velocities we use the Rotary Jib Standard constructed by UT Testing Centre and for higher air flow velocities we use the Wind Tunnel Standard that is obtained by the support of EU. We also calibrate thermometers that belong to anemometers. For more information you are welcome to contact Mr. Martin Vilbaste (contact data on our contact page).
Has been accredited
Measurement range |
Calibration and measurement capability |
Rotary Jib Standard (0.1…5) m/s |
(0.05…0.12) m/s |
Wind Tunnel Standard (0.5…20) m/s |
(0.08…0.6) m/s |
Calibration of air humidity measuring devices
|
price according to agreement |
For calibrating air humidity measuring devices we use the secondary humidity standard obtained by the support of EU. The main parts of the standard are the climatic chamber and the chilled mirror dew-point hygrometer. We also calibrate air thermometers that belong to humidity measuring devices. Estonian relative humidity calibration and measurement capabilities (CMC) have been published in the BIPM database. For more information you are welcome to contact Mr. Martin Vilbaste (contact data on our contact page).
Has been accredited
Measurement range |
Calibration and measurement capability |
Relative humidity |
at temperatures |
Reference standard |
Working standard |
(10…95) % |
(25…60) °C |
(0.7…2.3) % |
(0.9…2.4) % |
(25…95) % |
(20…25) °C |
(0.9…2.3) % |
(1.1…2.4) % |
(40…95) % |
(15…20) °C |
(1.2…2.3) % |
(1.3…2.4) % |
(50…95) % |
(10…15) °C |
(1.5…2.3) % |
(1.6…2.4) % |
Calibration of air thermometers
|
price according to agreement |
For more information you are welcome to contact Mr. Martin Vilbaste (contact data on our contact page).
Has been accredited
Measurement range |
Calibration and measurement capability |
(-10…80) °C |
0.2 °C |
Calibration of differential manometers
|
price according to agreement |
For additional information you are welcome to contact Mr. Martin Vilbaste (contact data on our contact page).
Has been accredited
Measurement range |
Calibration and measurement capability |
(-1000…2000) Pa |
(1…5) Pa |
Calibration of nonautomatic electronic weighing instruments
|
price according to agreement |
For additional information you are welcome to contact Mr. Martin Vilbaste or Koit Herodes (contact data on our contact page).
Has been accredited
Measurement range |
Calibration and measurement capability |
(1…50) mg |
0.005 mg |
(50…500) mg |
0.01 mg |
(0.5…20) g |
0.03 mg |
(20…50) g |
0.04 mg |
(50…100) g |
0.07 mg |
(100…200) g |
0.09 mg |
(200…300) g |
0.16 mg |
(300…400) g |
0.18 mg |
(400…510) g |
0.3 mg |
Calibration of spectral irradiance sensors |
|
For additional information you are welcome to contact Riho Vendt (contact data on our contact page).
Has been accredited
Measurement range |
Calibration and measurement capability |
100 nW/(m2 nm)…500 mW/(m2 nm) |
|
(290…360) nm |
(2.2…10) % of measured value |
(360…900) nm |
(1.1…2.2) % of measured value |
Calibration of spectral radiance sensors |
|
For additional information you are welcome to contact Riho Vendt (contact data on our contact page).
Has been accredited
Measurement range |
Calibration and measurement capability |
50 nW/(m2 sr nm)…100 mW/(m2 sr nm) |
|
(290…360) nm |
(3.6…16) % of measured value |
(360…900) nm |
(1.4…3.6) % of measured value |
For additional information you are welcome to contact Riho Vendt (contact data on our contact page).
Has been accredited
Measurement range |
Calibration and measurement capability |
100 nW/(m2 nm)…500 mW/(m2 nm) |
|
(340…900) nm |
(1.4…3.0) % of measured value |
For additional information you are welcome to contact Riho Vendt (contact data on our contact page).
Has been accredited
Measurement range |
Calibration and measurement capability |
Regular transmittance 0.0001…1 |
|
(300…900) nm |
(0.4…10) % |