{"id":6,"date":"2024-04-04T04:31:15","date_gmt":"2024-04-04T01:31:15","guid":{"rendered":"https:\/\/sisu.ut.ee\/heritage-analysis\/62-mass-spectrometry\/"},"modified":"2024-07-18T14:32:04","modified_gmt":"2024-07-18T11:32:04","slug":"62-mass-spectrometry","status":"publish","type":"page","link":"https:\/\/sisu.ut.ee\/heritage-analysis\/62-mass-spectrometry\/","title":{"rendered":"6.2. Mass spectrometry"},"content":{"rendered":"<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p>This section summarises practical analysis aspects of mass spectrometry, and guidelines on choosing a suitable ion source and mass analyser combination for investigating different materials are given.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\"><span style=\"color: #b22222;\"><span style=\"line-height: normal;\"><b><span lang=\"EN-GB\">Practical aspects of direct MS<\/span><\/b><\/span><\/span><\/h2>\n\n\n\n<p>Mass spectrometry is a powerful technique that provides valuable molecular information about the components of complex materials. It is often combined with <strong>gas chromatography (GC)<\/strong> or <strong>liquid chromatography (LC)<\/strong>. It is also applicable as a<strong> stand-alone (direct) technique<\/strong> allowing to combine different <strong>ionisation methods<\/strong> (electrospray ionisation (ESI), atmospheric pressure chemical ionisation (APCI), matrix-assisted laser desorption ionisation (MALDI), etc.) and<strong> mass analysers<\/strong> (quadrupole (Q), ion trap (IT), Fourier transform ion cyclotron resonance (FT-ICR), time of flight (ToF), etc.).<\/p>\n\n\n\n<div class=\"wp-block-group is-vertical is-layout-flex wp-container-core-group-is-layout-8cf370e7 wp-block-group-is-layout-flex\">\n<p><strong>Direct MS<\/strong> can be very useful in the analysis of cultural heritage samples. The main advantages of direct MS compared to LC-MS and GC -MS analysis are:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li><strong style=\"font-size: revert; text-align: var(--bs-body-text-align);\">Possibility to use smaller sample amounts<\/strong><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\">: about a few \u00b5g or even less for direct MS;<\/span><\/li>\n<\/ul>\n\n\n\n<div style=\"height:0px\" aria-hidden=\"true\" class=\"wp-block-spacer wp-container-content-9760934e\"><\/div>\n\n\n\n<ul class=\"wp-block-list\">\n<li> <strong style=\"font-size: revert; text-align: var(--bs-body-text-align);\">Simpler sample preparation:<\/strong><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\"> possible to use a wider selection of solvents (dichloromethane, acetone, methanol, etc.), no special sample pre-treatment (derivatisation, etc.) needed and no limitations that could come from the column or chromatographic system;<\/span><\/li>\n<\/ul>\n\n\n\n<div style=\"height:0px\" aria-hidden=\"true\" class=\"wp-block-spacer wp-container-content-9760934e\"><\/div>\n\n\n\n<ul class=\"wp-block-list\">\n<li><p><strong style=\"font-size: revert; text-align: var(--bs-body-text-align);\">Possible to perform analysis directly<\/strong><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\"> (i.e., without dissolution, extraction) from the <\/span><strong style=\"font-size: revert; text-align: var(--bs-body-text-align);\">solid samples<\/strong><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\"> (like aged adhesive residues, dyes from textile fibres, etc.). This kind of analysis can be made with <\/span><strong style=\"font-size: revert; text-align: var(--bs-body-text-align);\">MALDI-MS<\/strong><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\">, direct temperature MS techniques, such as <\/span><strong style=\"font-size: revert; text-align: var(--bs-body-text-align);\">Direct Inlet-Mass spectrometry (DI-MS)<\/strong><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\">, <\/span><strong style=\"font-size: revert; text-align: var(--bs-body-text-align);\">Direct Exposure-Mass Spectrometry (DE-MS)<\/strong><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\">, and<\/span><strong style=\"font-size: revert; text-align: var(--bs-body-text-align);\"> Direct Temperature Resolved-Mass Spectrometry (DTMS)<\/strong><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\">.[1]<\/span><span style=\"font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\"> <\/span><\/p><\/li>\n<\/ul>\n<\/div>\n\n\n\n<div class=\"wp-block-group is-vertical is-layout-flex wp-container-core-group-is-layout-8cf370e7 wp-block-group-is-layout-flex\">\n<p>The list of different MS instruments is long. Essentially, all the ion sources can be combined with the available mass analysers, but some combinations have proven more efficient and easier to manage. E.g., the MALDI source is very suitable for ToF, while ESI and APCI work well with QQQ. Another prominent class of direct MS techniques are the direct temperature MS techniques, like DTMS, DI-MS, and DE-MS, which use high temperature to desorb\/pyrolyze the analysed compounds. These techniques use the EI\/CI source (also widely used for GC-MS).<\/p>\n\n\n\n<p>From the practical aspect, the choice of the ion source is very important, and it is mainly determined by the type and properties of the sample. Based on the sample, the following aspects should be considered: (1) in which solvents (if at all) and how well the sample dissolves, (2) whether it ionizes better in the positive or negative mode, (3) whether the material consists of small or larger molecules, (4) sample amount available, etc. For example, for ESI and APCI sources, the sample must be completely dissolved, but with MALDI also hazy solutions and even solid samples can be directly analysed.<\/p>\n\n\n\n<p><\/p>\n<\/div>\n\n\n\n<div class=\"wp-block-group is-vertical is-layout-flex wp-container-core-group-is-layout-8cf370e7 wp-block-group-is-layout-flex\">\n<p>The overall steps of direct MS analysis are the following (see more information in Table 1):<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li><strong>Sample preparation<\/strong> (depends greatly on the instrument, ion source): for ESI\/APCI-MS and also for MS with EI\/CI the sample must be dissolved\/extracted. Filtration, sometimes dilution, rarely derivatisation may be needed. For MALDI-MS and also for high-temperature direct MS techniques, the sample can be solid or partially dissolved.\n<ul class=\"wp-block-list\">\n<li><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\">For the analysis with <\/span><strong style=\"font-size: revert; text-align: var(--bs-body-text-align);\">HRMS instruments<\/strong><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\">, calibration standards should be used to calibrate the <\/span><em style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\">m\/z<\/em><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\"> axes of the measured MS spectra to obtain the highest <\/span><em style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\">m\/z<\/em><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\"> accuracy possible (e.g., <\/span><em style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\">m\/z<\/em><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\"> errors less than 2 ppm) [2-5].\u00a0For this, a solution of calibration standards, also named internal standards (at least 3 compounds), is added to the sample during the preparation (usually calibration solution is mixed with sample solution).\u00a0<\/span> <\/li>\n<\/ul>\n<\/li>\n<\/ul>\n\n\n\n<div style=\"height:0px\" aria-hidden=\"true\" class=\"wp-block-spacer wp-container-content-9760934e\"><\/div>\n\n\n\n<ul class=\"wp-block-list\">\n<li><strong>Measurement of the sample with MS:<\/strong>\n<ul class=\"wp-block-list\">\n<li><strong>Introduction of the sample into the ion source.<\/strong> Usually, a sample solution is introduced into the ion source with a micro syringe (with ESI-, APCI-MS), using a special target plate (with MALDI-MS) or direct inlet probe (with DTMS, DI-MS and DE-MS).<\/li>\n\n\n\n<li><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\">For <\/span><strong style=\"font-size: revert; text-align: var(--bs-body-text-align);\">external calibration of the HRMS instrument<\/strong><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\"> (to verify the overall conditions of the HRMS instrument), the\u00a0<\/span><strong style=\"font-size: revert; text-align: var(--bs-body-text-align);\">calibration standard solution<\/strong><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\"> should be analysed similarly as the sample.<\/span> <\/li>\n\n\n\n<li><strong style=\"font-size: revert; text-align: var(--bs-body-text-align);\">Optimisation of the parameters of ion source and mass analyser.<\/strong><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\"> For that, special software of the instrument is used. Usually, it is done once during method development and then adjusted if needed. If all the parameters are fixed, then the mass spectrum of the sample is recorded.\u00a0<\/span><\/li>\n<\/ul>\n<\/li>\n<\/ul>\n\n\n\n<div style=\"height:0px\" aria-hidden=\"true\" class=\"wp-block-spacer wp-container-content-9760934e\"><\/div>\n\n\n\n<ul class=\"wp-block-list\">\n<li><strong style=\"font-size: revert; text-align: var(--bs-body-text-align);\">Interpretation of results.<\/strong><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\"> Interpretation of mass spectrum is based on <\/span><em style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\">m\/z<\/em><span style=\"font-size: revert; font-weight: var(--bs-body-font-weight); text-align: var(--bs-body-text-align);\"> values and the corresponding ion formulas. See more details below.<\/span> <\/li>\n<\/ul>\n<\/div>\n\n\n\n<p><\/p>\n\n\n\n<p><span style=\"line-height: normal;\"><span lang=\"EN-GB\">Please see Table 1 and the videos below for more detailed information.<\/span><\/span> The information in Table 1 is a very broad generalisation. Exceptions can be found in almost all statements presented in the table.<\/p>\n\n\n\n<p><strong>Table 1.<\/strong> Comparison of some mass spectrometric techniques based on the ion source and their suitability for the analysis of cultural heritage materials. Under the ion source (in bold), in brackets are the more common mass analysers that are combined with the source. [1,6]<\/p>\n\n\n\n<figure class=\"wp-block-table has-small-font-size\"><table class=\"table table-hover\"><thead><tr><th class=\"has-text-align-left\" data-align=\"left\"><\/th><th class=\"has-text-align-left\" data-align=\"left\"><strong>EI and CI  <\/strong>(+ Q, IT)<\/th><th class=\"has-text-align-left\" data-align=\"left\"><strong>MALDI <\/strong>(+ToF, FT-ICR, also Q, QQQ, IT and combinations)<\/th><th class=\"has-text-align-left\" data-align=\"left\"><strong>ESI and APCI <\/strong>(+QQQ, IT, FT-OT, FT-ICR, also ToF and combinations)<\/th><\/tr><\/thead><tbody><tr><td class=\"has-text-align-left\" data-align=\"left\"><strong>Sample preparation<\/strong><\/td><td class=\"has-text-align-left\" data-align=\"left\">\u2212 <strong>Gas-phase ion source<\/strong> \u2192 analytes are introduced in gas phase.<br>\u2212 Mainly used<br>(a) in hyphenation with gas chromatography, as detector;<br>(b) with<strong> direct temperature MS techniques<\/strong>: dissolved\/extracted or solid sample is inserted into the ion source (through a specific inlet) and gradually heated \u2192 compounds are desorbed\/pyrolyzed.<br>\u2212 <strong>Suitable solvents:<\/strong> methanol, ethanol, toluene, DCM, etc.<\/td><td class=\"has-text-align-left\" data-align=\"left\">\u2212 The sample is co-crystallized with a matrix material: sample (solution) is mixed with matrix solution and placed on MALDI target plate, solvent(s) is (are) evaporated, and dried crystals are analysed.<br>\u2212 The sample does not need to be fully soluble (the solutions can be hazy).<br>\u2212<strong> Matrix substances:<\/strong><br><strong>*positive ion mode:<\/strong> DHB, CHCA, SA, etc.<br><strong>*negative ion mode:<\/strong> 9-AA, 3-AA, etc.<br>\u2212 Suitable solvents: water, methanol, toluene, DCM, DMSO, etc.<\/td><td class=\"has-text-align-left\" data-align=\"left\">\u2212 <strong>Liquid phase ion source<\/strong> \u2192 analytes are introduced in liquid phase, most often in combination with liquid chromatography in MS detector.<br>\u2212 Sample is <strong>dissolved\/extracted<\/strong> with solvent(s). Sample solution must be filtered (or centrifuged)!<br>\u2212 <strong>Suitable solvents:<\/strong><br><strong>ESI:<\/strong> most commonly methanol, acetonitrile, water, and their mixtures; less often toluene, ethyl acetate, dichloromethane, tetrahydrofuran, etc.<br><strong>APCI:<\/strong> methanol, hexane, acetonitrile, dichloromethane, etc.<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\"><strong>Sample size<\/strong><\/td><td class=\"has-text-align-left\" data-align=\"left\"><strong>Less than 1 \u00b5g,<\/strong> usually only few ng of a compound in a mixture.<\/td><td class=\"has-text-align-left\" data-align=\"left\"><strong>Less than 1 \u00b5g <\/strong>(<strong>per mL<\/strong> of sample solution).<\/td><td class=\"has-text-align-left\" data-align=\"left\"><strong>ESI <\/strong>and<strong> APCI:<\/strong> approx. <strong>1 mg<\/strong> <strong>per mL<\/strong> of sample solution.<br><strong>Nano-ESI:<\/strong> <strong>few dozen \u00b5g per mL<\/strong> of sample solution.<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\"><strong>Material types that can be analysed<\/strong><\/td><td class=\"has-text-align-left\" data-align=\"left\">Compounds with sufficient vapor pressure (boiling points below ca 600 \u00b0C) directly; non-volatile compounds, such as resinous materials (resins, varnishes, tars), oils\/fats, waxes, saccharides, proteins\/peptides, dyes, polymers can be analysed using derivatization and\/or pyrolysis.<\/td><td class=\"has-text-align-left\" data-align=\"left\">Resinous materials (resins, varnishes, tars), oils\/fats, waxes, saccharides, proteins\/peptides, dyes, polymers, inorganic compounds.<\/td><td class=\"has-text-align-left\" data-align=\"left\"><strong>ESI:<\/strong> Resinous materials (resins, varnishes, tars), oils\/fats, waxes, saccharides, proteins\/peptides, dyes, polymers.<br><strong>APCI:<\/strong> Resinous materials (resins, varnishes, tars), oils\/fats, waxes, saccharides, dyes.<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\"><strong>Information in mass spectra<\/strong><br>(more info)<\/td><td class=\"has-text-align-left\" data-align=\"left\"><strong>EI:<\/strong><br>\u2212 <strong>Highly fragmenting <\/strong>\u2192 fragment ions are detected, sometimes also the molecular ion \u2192 characteristic fragmentation pattern for compounds.<br>\u2212 <strong>Only positive ions<\/strong> are generated.<br><strong>CI:<\/strong><br>\u2212<strong> Less fragmenting<\/strong> \u2192 molecular ions and fragment ions \u2192 simple mass spectrum<br>\u2212 Adduct ions with the reagent gas may form.<br>\u2212 <strong>Positive and negative ions<\/strong> can be generated.<\/td><td class=\"has-text-align-left\" data-align=\"left\">\u2212 <strong>Positive and negative ions<\/strong> can be generated \u2192 mainly adduct ions with H+, Na+, K+, Cl\u2013, etc. \u2192 characteristic to specific materials.<br>\u2212 Mainly singly charged ions.<br>\u2212 Low fragmentation.<br>\u2212 Macromolecules can be ionised.<\/td><td class=\"has-text-align-left\" data-align=\"left\">\u2212 <strong>Positive and negative ions.<\/strong><br>\u2212 Low fragmentation.<br><strong>ESI:<\/strong><br>\u2212 Singly and multiply charged ions \u2192 multiply charged ions: analysis of compounds with very high mass (approx. 100000 Da), singly charged ions: information about small molecules.<br>\u2212 Generates mainly adduct ions with H+, Na+, K+, Cl\u2013, etc.<br>\u2212 Preferably <strong>more polar compounds<\/strong> are ionised.<br><strong>APCI:<\/strong><br>\u2212 Mainly singly charged ions.<br>\u2212 Mainly protonated or deprotonated ions.<br>\u2212 <strong>Preferably less polar to nonpolar small<\/strong> (up to 2000 Da) <strong>compounds<\/strong> are ionised.<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\"><strong>Advantages<\/strong><\/td><td class=\"has-text-align-left\" data-align=\"left\">\u2212<strong> Very small sample can be analysed<\/strong> \u2192 very sensitive.<br>\u2212 Very good ionisation efficiency.<br>\u2212 Fragmentation leads to characteristic spectra (\u201cfingerprints\u201d) and gives information about the structure of the compounds.<br>\u2212 Extensive and well accessible spectral libraries exist that are help with identification.<br>\u2212 Qualitative and quantitative analysis.<\/td><td class=\"has-text-align-left\" data-align=\"left\">\u2212 <strong>Can be used for partially dissolved samples <\/strong>(hazy solution) or<strong> directly on the solid samples<\/strong> without extraction\/decomposing<br>\u2212 Easy sample preparation.<br>\u2212 Aged, oxidised, degraded samples can be analysed, incl. materials that are (partly) polymeric and non-volatile.<br>\u2212 No extensive fragmentation \u2192 typically ions corresponding to original compounds and their derivatives are detected<br>\u2212 Good ionization efficiency.<\/td><td class=\"has-text-align-left\" data-align=\"left\">\u2212 <strong>Soft ionisation.<\/strong><br>Qualitative and quantitative analysis.<br><strong>ESI:<\/strong><br>\u2212 Suitable for the analysis of most organic compounds.<br>\u2212 Suitable for analytes with low concentration (in case of nano-ESI).<br><strong>APCI:<\/strong><br>\u2212 Less matrix effects compared to ESI.<\/td><\/tr><tr><td class=\"has-text-align-left\" data-align=\"left\"><strong>Disadvantages<\/strong><\/td><td class=\"has-text-align-left\" data-align=\"left\">\u2212 <strong>Molecules can be extensively fragmented<\/strong> (e.g., aliphatic compounds) and spectra may be insufficiently characteristic for identification. In particular, molecular ions are not always observed.<br>\u2212 Instrument needs constant monitoring and maintenance.<br>\u2212 Skilled operator is needed.<br>\u2212 CI \u2192 limited amount of structural information.<\/td><td class=\"has-text-align-left\" data-align=\"left\">\u2212 <strong>Selection of matrix substance and suitable solvents is crucial <\/strong>\u2192 the matrix and solvent(s) must be compatible with sample.<br>\u2212 Poorly reproducible spectra.<br>\u2212 Not suitable for quantitative analysis.<br>\u2212 Interpretation relies on comparison to standard materials and literature.<br>\u2212 Instrument needs constant monitoring and maintenance.<br>\u2212 Skilled operator is needed.<\/td><td class=\"has-text-align-left\" data-align=\"left\">\u2212 Liquid environment for the ionisation needed \u2192 <strong>Sample has to be dissolved <\/strong>(only the dissolved\/extracted part of the sample can be analysed).<br>\u2212 Higher sample amount needed.<br>\u2212 Instrument needs constant monitoring and maintenance.<br>\u2212 Skilled operator is needed.<br><strong>ESI:<\/strong><br>\u2212 <strong>Matrix effects<\/strong><br>\u2212 Spray stability depends on the solvent.<br><strong>APCI:<\/strong><br>\u2212 Only suitable for small molecules (under 2000 Da).<br>\u2212 System needs constant cleaning \u2192 contamination of the corona needle and MS inlet.<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p><\/p>\n\n\n\n<p><span style=\"line-height: normal;\"><span lang=\"EN-GB\">As demonstrated in Table 1, the information obtained with different MS instruments can vary because the ionisation method is different. As an example, in Fig. 1, mass spectra of <strong>shellac resin<\/strong>, <strong>colophony resin<\/strong> (both in positive mode) and <strong>extract of dyer\u2019s madder plan<\/strong>t (negative mode) obtained with <strong>FT-ICR-MS<\/strong> coupled to <strong>MALDI<\/strong>, <strong>APCI <\/strong>and <strong>ESI<\/strong> sources, respectively are presented. The selection of ion mode \u2013 positive or negative depends on the analysed material (e.g., natural resins preferably give positive ions while dyes give negative ions). In positive mode, ESI and APCI give mostly protonated ions, sometimes also adducts with Na<sup>+<\/sup>. With MALDI, preferably adducts with Na<sup>+<\/sup>, sometimes also K<sup>+<\/sup>, Al<sup>+<\/sup>, etc., are observed (such adducts are common for compounds with a higher amount of oxygen in them). In negative ion mode, mainly deprotonated ions are observed but also the addition of Cl<sup>\u2013<\/sup> or other anions may be possible (see Fig.1c). With ESI, more polar compounds can be detected (reins, dyes, oils\/fats, etc.), MALDI and APCI are also suitable for less polar compounds (waxes, hydrocarbons). <\/span><\/span><\/p>\n\n\n\n<figure class=\"wp-block-image aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"1107\" height=\"1476\" src=\"https:\/\/sisu.ut.ee\/wp-content\/uploads\/sites\/285\/6.2_Fig1_MS_2.png\" alt=\"\" class=\"wp-image-1138\" srcset=\"https:\/\/sisu.ut.ee\/wp-content\/uploads\/sites\/285\/6.2_Fig1_MS_2.png 1107w, https:\/\/sisu.ut.ee\/wp-content\/uploads\/sites\/285\/6.2_Fig1_MS_2-225x300.png 225w, https:\/\/sisu.ut.ee\/wp-content\/uploads\/sites\/285\/6.2_Fig1_MS_2-768x1024.png 768w\" sizes=\"auto, (max-width: 1107px) 100vw, 1107px\"><figcaption class=\"wp-element-caption\">Fig. 1. Mass spectra of a) shellac resin (obtained with MALDI(+)-FT-ICR-MS); b) colophony resin (obtained with APCI(+)-FT-ICR-MS); c) dyer\u2019s madder plant [7] (obtained with ESI(-)-FT-ICR-MS). CS denotes ions corresponding to protonated calibration standards (only the visible CS are highlighted).<\/figcaption><\/figure>\n\n\n\n<p><span style=\"line-height: normal;\"><span lang=\"EN-GB\"><strong>Interpretation of the obtained mass spectra<\/strong> is based on the identification of detected <strong>ions\/fragments<\/strong> and also on the <strong>isotope pattern<\/strong>, which helps to determine the charge of the ion (by the spacing between the satellite peaks), as well as the presence of some elements with characteristic isotope patterns (such as Cl, Br, S). If a <strong>HRMS instrument<\/strong> is used, like FT-ICR-MS, Orbitrap, or high-end ToF, then the obtained <strong><i>m\/z<\/i> values<\/strong> are <strong>very accurate<\/strong> (four to six decimals) and often a single mass analyser is enough for assigning correct ion formulas and, hence<\/span><\/span>, <span style=\"line-height: normal;\"><span lang=\"EN-GB\">identifying the corresponding compounds, as demonstrated on Fig.1 (reference mass spectra are still very helpful). With <strong>LRMS instruments<\/strong> only <strong>nominal <i>m\/z<\/i> values<\/strong> are obtained. This makes <i>m\/z<\/i>-based identifications difficult or impossible because the number of potential ion formulas corresponding to a nominal value can be very high. In the case of LRMS, tandem instruments (see section 5.2) are helpful. This technique allows to select specific <i>m\/z<\/i> values, fragment them, and analyse the obtained fragments. In this case, identification is mostly based on fragmentation patterns \u2013 different compounds (even with the same nominal <i>m\/z<\/i> value) can fragment differently.<\/span><\/span><\/p>\n\n\n\n<h2 class=\"wp-block-heading\"><em><span style=\"color: #b22222;\">Analysis with MALDI-, ESI- and APCI-FT-ICR-MS<\/span><\/em><\/h2>\n\n\n\n<p>In the following video, Dr Anu Teearu-Ojak\u00e4\u00e4r introduces the principles of MALDI-FT-ICR-MS, sample preparation and how to perform analysis with this instrument.<\/p>\n\n\n\n<p><\/p><div class=\"ratio ratio-16x9 mb-3\"><div class=\"video-placeholder-wrapper video-placeholder-wrapper--16x9\">\n\t\t\t    <div class=\"video-placeholder d-flex justify-content-center align-items-center\">\n\t\t\t        <div class=\"overlay text-white p-2 w-100 text-center d-block justify-content-center align-items-center\">\n\t\t\t            <div>To view third-party content, please accept cookies.<\/div>\n\t\t\t            <button class=\"btn btn-secondary btn-sm mt-1 consent-change\">Change consent<\/button>\n\t\t\t        <\/div>\n\t\t\t    <\/div>\n\t\t\t<\/div>\n<\/div>\n\n\n\n<p>In the following video, Dr Anu Teearu-Ojak\u00e4\u00e4r introduces the FT-ICR-MS with APCI and ESI ionisation sources, sample preparation and how to perform analysis with ESI-FT-ICR-MS.<\/p>\n\n\n\n<p><\/p><div class=\"ratio ratio-16x9 mb-3\"><div class=\"video-placeholder-wrapper video-placeholder-wrapper--16x9\">\n\t\t\t    <div class=\"video-placeholder d-flex justify-content-center align-items-center\">\n\t\t\t        <div class=\"overlay text-white p-2 w-100 text-center d-block justify-content-center align-items-center\">\n\t\t\t            <div>To view third-party content, please accept cookies.<\/div>\n\t\t\t            <button class=\"btn btn-secondary btn-sm mt-1 consent-change\">Change consent<\/button>\n\t\t\t        <\/div>\n\t\t\t    <\/div>\n\t\t\t<\/div>\n<\/div>\n\n\n<p><\/p><div class=\"accordion mb-3\">\n        <div class=\"accordion-item accordion-item--white\">\n        <h2 class=\"accordion-header\" id=\"accordion-69d9513476bc9-heading\">\n            <button class=\"accordion-button collapsed\" type=\"button\" data-bs-toggle=\"collapse\" data-bs-target=\"#accordion-69d9513476bc9-collapse\" aria-expanded=\"true\" aria-controls=\"accordion-69d9513476bc9-collapse\"><span style=\"font-size: 16px;\"><strong>References<\/strong> (click here)<\/span><\/button>\n        <\/h2>\n        <div id=\"accordion-69d9513476bc9-collapse\" class=\"accordion-collapse collapse\" aria-labelledby=\"accordion-69d9513476bc9-heading\">\n            <div class=\"accordion-body\">\n<ol>\n<li>Colombini, M. P.; Modugno, F. <em>Organic Mass Spectrometry in Art and Archaeology<\/em>; John Wiley &amp; Sons, Ltd.: Chichester, UK, 2009.<\/li>\n<li>Teearu, A.<em> Development of MALDI-FT-ICR-MS Methodology for the Analysis of Resinous Materials<\/em>, PhD thesis. University of Tartu Press: Tartu, Estonia, 2017.<\/li>\n<li>Teearu, A.; Vahur, S.; Rodima, T.; Herodes, K.; Bonrath, W.; Netscher, T.; Tshepelevitsh, S.; Trummal, A.; L\u00f5kov, M.; Leito, I. Method Development for the Analysis of Resinous Materials with MALDI-FT-ICR-MS: Novel Internal Standards and a New Matrix Material for Negative Ion Mode. <em>J. Mass. Spectrom.<\/em> <strong>2017<\/strong>, <em>52<\/em> (9), 603\u2013617. https:\/\/doi.org\/10.1002\/jms.3943.<\/li>\n<li>Teearu, A.; Vahur, S.; Haljasorg, U.; Leito, I.; Haljasorg, T.; Toom, L. 2,5-Dihydroxybenzoic Acid Solution in MALDI-MS: Ageing and Use for Mass Calibration. <em>J. Mass. Spectrom<\/em>. <strong>2014<\/strong>, <em>49 <\/em>(10), 970\u2013979. https:\/\/doi.org\/10.1002\/jms.3395.<\/li>\n<li>McIver, R. T.; McIver, J. R. <em>Fourier Transform Mass Spectrometry<\/em>; IonSpec Corporation: Lake Forest, CA, USA, 2006.<\/li>\n<li>Stuart, B. H. <em>Analytical Techniques in Materials Conservation<\/em>; John Wiley &amp; Sons, Ltd.: Chichester, UK, 2007.<\/li>\n<\/ol>\n<p><\/p><\/div>\n        <\/div>\n        <\/div>\n    <\/div>\n\n\n\n<p>\u00a0<\/p>\n","protected":false},"excerpt":{"rendered":"<p>This section summarises practical analysis aspects of mass spectrometry, and guidelines on choosing a suitable ion source and mass analyser combination for investigating different materials are given. Practical aspects of direct MS Mass spectrometry is a powerful technique that provides &#8230;<\/p>\n","protected":false},"author":151,"featured_media":0,"parent":0,"menu_order":0,"comment_status":"closed","ping_status":"closed","template":"","meta":{"_acf_changed":false,"inline_featured_image":false,"footnotes":""},"class_list":["post-6","page","type-page","status-publish","hentry"],"acf":[],"_links":{"self":[{"href":"https:\/\/sisu.ut.ee\/heritage-analysis\/wp-json\/wp\/v2\/pages\/6","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/sisu.ut.ee\/heritage-analysis\/wp-json\/wp\/v2\/pages"}],"about":[{"href":"https:\/\/sisu.ut.ee\/heritage-analysis\/wp-json\/wp\/v2\/types\/page"}],"author":[{"embeddable":true,"href":"https:\/\/sisu.ut.ee\/heritage-analysis\/wp-json\/wp\/v2\/users\/151"}],"replies":[{"embeddable":true,"href":"https:\/\/sisu.ut.ee\/heritage-analysis\/wp-json\/wp\/v2\/comments?post=6"}],"version-history":[{"count":21,"href":"https:\/\/sisu.ut.ee\/heritage-analysis\/wp-json\/wp\/v2\/pages\/6\/revisions"}],"predecessor-version":[{"id":1144,"href":"https:\/\/sisu.ut.ee\/heritage-analysis\/wp-json\/wp\/v2\/pages\/6\/revisions\/1144"}],"wp:attachment":[{"href":"https:\/\/sisu.ut.ee\/heritage-analysis\/wp-json\/wp\/v2\/media?parent=6"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}